► Temperature Range : −200 °C to 1200 °C
► Temperature Accuracy : 0.008 °C at 0 °C
► Display resolution : 6 1/2 digits
► Resolution : 0.001 °C
► 다채널 구성 가능
1586A Super-DAQ Precision Temperature Scanner
The 1586A Super-DAQ collects time-stamped precision temperature and electrical measurements for data analysis by technicians, engineers,
and quality control personnel to verify process control, analyze interactive systems, ensure conformance to quality standards,
or to correlate related events for R&D or troubleshooting. Measurement data and statistics can be viewed in tabular format for all active channels.
With the graphing feature, up to four channels can be plotted at the same time, making it easy to quickly assess test setup and results before analyzing the data on a PC.
When configured with the DAQ-STAQ Multiplexer, the Super-DAQ has the accuracy of the best benchtop reference thermometer readouts for calibration of PRTs, RTDs, thermistors, or thermocouples.
Lab efficiency can be increased when the Super-DAQ is connected to a Fluke Calibration drywell or bath and Automated Sensor Test routines are run.
The 1586 is ideal for a number of applications such as thermal mapping, temperature validation, process sensor calibration, and more.
These applications are found in a various industries including pharmaceutical, biotechnology, food processing, aerospace, and automotive.
Some temperature data acquisition applications require compliance to government and standards agency requirements.
TQSoft and TQAero thermal validation software help engineers, technicians and lab managers to meet the key guidelines for temperature data collection.
These software solutions provide a complete qualification, reporting, documentation, and audit trail management system.
General Specifications |
|
Mains Voltage | 100 V Setting: 90 V to 110 V 120 V Setting: 108 V to 132 V 220 V Setting: 198 V to 242 V 240 V Setting: 216 V to 264 V |
Frequency | 47 Hz to 440 Hz |
Power Consumption | 36 VA peak (24 W average) |
Environment Temperature | Operating: 0 °C to 50 °C Full accuracy: 18 °C to 28 °C Storage: −20 °C to 70 °C Warm-up: 1 hour to full accuracy specifications |
Relative Humidity (non-condensing) | Operating: 0 °C to 30 °C <80 %; 30 °C to 50 °C <50 % Storage: −20 °C to 70 °C <95 % |
Altitude | Operating: 2,000 m Storage: 12,000 m |
Vibration and Shock | Complies with MIL-PRF-28800F Class 3 |
Channel Capacity | Total analog channels: 45 Voltage/resistance channels: 41 Current channels: 5 Digital I/O: 8 bits Totalizer: 1 Alarm outputs: 6 Trigger input: 1 |
Channel Input Connector Types | 1586A Channel 001 (front panel): gold pins, bare wire, spade lugs, banana plugs DAQ-STAQ Multiplexer: gold pins, bare wire, mini spade lugs, mini banana plugs, mini-jack thermocouple High-Capacity Module: gold pins, bare wire |
Input Protection | 50 V all functions, terminals and ranges |
Math Channels | Number of channels: 20 Operations: sum, difference, multiply, divide, polynomial, power, square root, reciprocal, exponential, logarithm, absolute value, average, maximum, minimum |
Triggers | interval, external (trigger input), alarm, remote (bus), manual, automated test |
Memory | Scan data RAM: 75,000 readings with timestamp Data/Setup flash memory: 20 MB |
USB Host Port | Connector type: Type A Function: Memory File system: FAT32 Memory capacity: 32 GB |
USB Device Port | Connector type: Type B Class: Instrument Function: Control and data transfer Command protocol: SCPI |
LAN | Function: Control and data transfer Network protocols: Ethernet 10/100, TCP/IP Command protocol: SCPI |
RS-232 | Connector: D-sub 9 pin (DE-9) Baud rates: 1200, 2400, 4800, 9600, 19200, 38400 Function: Temperature source control output |
Dimensions | Height: 150 mm Width: 245 mm Depth: 385 mm Weight: 6 kg (typical configuration) Shipping Weight: 9.5 kg (typical configuration) |
Conformity | CE, CSA, IEC 61010 3rd ed. |