제품소개 Products



tanδ(Dissipation Factor)Standard TYPE DDF-A Series 모델명: TYPE DDF-A Series
▶ DDF-A can be directly connected to 4-Terminal LCR Meter for calibration due to 4-Terminal BNC structure.
▶ tanδ(Dissipation Factor)calibration at 120Hz can be made.
▶ Suitable for calibration standard due to excellent accuracy, temperature coefficient, stability.
▶ Switchable to arbitrary tanδ (Dissipation Factor) removing/inserting a SHORT plug.
▶ Custom value tanδ(Dissipation Factor)is available.
제품 카달로그
견적문의
DDF-A Series are 4-Terminal variable tanδ Standards. Directly connected to LCR Meter due to BNC structure.
Highly stable tanδ(Dissipation Factor)standard which is made of ceramic capacitor and accurate resistor.
Item/Model |
TYPE DDF100A |
TYPE DDF25A |
Used Frequency |
120 Hz |
|
Nominal tanδ(%) |
0.01%,0.10%,0.50% Total 3 values |
0.05%,0.15%,0.25% Total 3 values |
tanδ Accuracy |
Converted to set tanδ value Less Than ±1 % |
|
Capacitance |
13.263 μF |
|
Capacitance Tolerance |
±0.1 % |
|
Temperature Coefficient |
±50 ppm/℃ |
|
Stability |
±100 ppm/Year |
|
Max. Applied Voltage/Current |
Less Than 50 Vdc(15 Vac)/ 1A |
|
Dimension(W×D×H)mm |
140mm×100mm×90mm(Including SHORT Plug and rubber feet) |
|
Weight |
Approx.1 kg |
-
Specifications are subject to change without notice due to improvement
-
SunJEM will design products to meet your individual requirements.