▶ Residual oil measurement of the vaporous oil content from 0.001...5 mg/m³, 3...9 bar.
▶ High-precision PID sensor, innovative „Forced Pressure Variation“ measuring method, with integrated display,
▶ 4...20 mA analog output and digital Modbus RTU interface, incl. calibration certificate, in a robust flight case. Connection cable ODU/ODU 5m
Process reliability:
All functions / components are monitored internally. A complete function test report can be printed out using the service software.
On-site calibration:
Calibration can be carried out in the field using test gas cylinders. The service software can be used to generate a
verification report (as-found data) and a calibration report (as-left data) can be generated.
Ideal for Portable Measuring:
Compact Measuring devices, easy sampling and quickly ready for Measuring
Simple installation
Air inlet / air outlet - sampling via PTFE hose
or stainless steel line.
Digital and analogue interfaces as standard.
Connection via M12 plug
Service friendly - no downtime
Loosen four screws - replace sensor unit. At the touch of a button
a complete function test is carried out. This means that
Measuring can be continued almost without interruption.