► TM 6602 / 6612 / 6630
► Handheld thermometer for thermocouples and/or resistive probes
with memory
► TM thermometers are the perfect field tools for daily maintenance
and commissioning of temperature probes at 0.02% accuracy.
Easy-to-use and equipped with a large graphical display, TM thermometers are the perfect field tools for daily maintenance and commissioning of temperature probes. 3 models are available with 0.02% accuracy as follows:
TM 6602 for thermocouple measurement
TM 6612 for RTD measurement
TM 6630 for both thermocouple and RTD measurement
With a very low temperature coefficient (10 ppm/°C in thermocouples and 7ppm/°C in resistance), IP54 protection and robust design, TM thermometers are suitable for onsite use even in demanding environmental conditions. They are widely used in the energy field, engineering sector, metal industry and automotive field.
TM thermometers use a graphical interface making programming and reading easier, under graphical or test format. Due to memory feature (10,000 values) and extended functionalities (corrections of calibrated sensors, statistical functions…), TM are well adapted to different process job procedures and ensure full data traceability as well as advanced data exploitation. Use them together with DATACAL calibration software to display, manage data and issue your own customized calibration certificates.
TM thermometers are delivered with 4 alkaline AA batteries in standard. The optional battery charger used a rechargeable battery.
Key features:
High accuracy: 0.02% reading with an adjustable resolution of 1 µV (TM 6602 / 6630) and 1 mΩ (TM 6612 / 6630)
Values displayed in °C, °F, mV and Ω
Low temperature coefficient: 10 ppm /°C for thermocouples and 7 ppm/°C for resistance
Measurements with HOLD function
Correction of calibrated sensors
Display of minimum, maximum and average value
Backlight
10,000 values stored and displayed graphically
Specifications and performances of TM 6602 / 6630 @23°C ±5°C
DC voltage
Function | Range | Res. | Accuracy / 1 year | Measuring range |
IN | 100 mV | 1 µV | 0.020% RDG + 3 µV | -10 mV / 100 mV |
Temperature coefficient < 15 ppm/°C beyond reference domain
Thermocouples: Measurement
Type | Input range | Resolution | Accuracy / 1 year (Measurement) |
K | -250 to -200°C -200 to -120°C -120 to 0°C 0 to +1372°C |
0.2°C 0.1°C 0.05°C 0.05°C |
0.90°C 0.3°C 0.02% RDG + 0.12°C 0.02% RDG + 0.11°C |
T | -250 to -200°C -200 to -50°C -50 to +400°C |
0.2°C 0.05°C 0.05°C |
0.8°C 0.25°C 0.02% RDG + 0.09°C |
J | -210 to -200°C -200 to -120°C -120 to +60°C +60 to +1200°C |
0.05°C 0.05°C 0.05°C 0.05°C |
0.3°C 0.25°C 0.02% RDG + 0.11°C 0.02% RDG + 0.09°C |
E | -250 to -200°C -200 to -100°C -100 to +450°C +450 to 1000°C |
0.1°C 0.05°C 0.05°C 0.05°C |
0.55°C 0.2°C 0.02% RDG + 0.07°C 0.02% RDG + 0.05°C |
R | -50 to +150°C +150 to +550°C +550 to 1768°C |
0.5°C 0.2°C 0.1°C |
0.95°C 0.4°C 0.02% RDG + 0.3°C |
S | -50 to +150°C +150 to +550°C +550 to +1768°C |
0.5°C 0.2°C 0.1°C |
0.85°C 0.02% RDG + 0.4°C 0.02% RDG + 0.3°C |
B | +400 to +900°C +900 to +1820°C |
0.2°C 0.1°C |
0.95°C 0.5°C |
U | -200 to -100°C -100 to +600°C |
0.05°C 0.05°C |
0.35°C 0.2°C |
L | -200 to -100°C -100°C to +900°C |
0.05°C 0.05°C |
0.3°C 0.2°C |
C | -20 to +900°C +900 to 2310°C |
0.1°C 0.1°C |
0.3°C 0.02% RDG + 0.15°C |
N | -240 to -190°C -190 to -110°C -110 to +0°C +0 to +1300°C |
0.2°C 0.1°C 0.05°C 0.05°C |
0.6°C 0.25°C 0.15°C 0.02% RDG + 0.07°C |
Platine | -100 to +1400°C | 0.05°C | 0.3°C |
Mo | +0 to +1375°C | 0.05°C | 0.02% RDG + 0.1°C |
NiMo/NiCo | -50 to +1410°C | 0.05°C | 0.02% RDG + 0.35°C |
Thermocouples G, D: For specifications, refer to the instruction manual (Available on request)
Accuracy is given for reference @ 0°C.
When using the internal reference junction (except couple B) add an additional uncertainty of 0.3 °C at 0 °C.
It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered.
Temperature coefficient: <10% of accuracy /°C
Specifications and performances of TM 6612 / 6630 @23°C ±5°C
Resistance
Function | Range | Resolution | Accuracy / 1 year | Notes |
IN | 400 Ω | 1 mΩ | 0.012% RDG + 10 mΩ | Automatic detection: 2, 3 or 4 wires |
3600 Ω | 10 mΩ | 0.012% RDG + 100 mΩ | Automatic detection: 2, 3 or 4 wires |
Connection in resistance and RTDs through banana plugs or 4-pin round connector
Temperature coefficient: < 7 ppm/°C beyond reference domain
R internal: < 1 Ω
Noise VLF < 1 mV (@ F< 100 Hz)
Resistive probes: Measurement
Sensor | Range (Input and Output) | Resolution | Accuracy / 1 year (Measurement) |
Pt50 (= 3851) | -220°C to +850°C | 0.01°C | 0.012% RDG + 0.06°C |
Pt100 (= 3851) | -220°C to +850°C | 0.01°C | 0.012% RDG + 0.05°C |
Pt100 (= 3916) | -200°C to +510°C | 0.01°C | 0.012% RDG + 0.05°C |
Pt100 (= 3926) | -210°C to +850°C | 0.01°C | 0.012% RDG + 0.05°C |
Pt200 (= 3851) | -220°C to +120°C | 0.01°C | 0.012% RDG + 0.12°C |
Pt500 (= 3851) | -220°C to +1200°C | 0.01°C | 0.012% RDG + 0.07°C |
Pt1000 (= 3851) | -220°C to +760°C | 0.01°C | 0.012% RDG + 0.05°C |
Ni100 (= 618) | -60°C to 180°C | 0.01°C | 0.012% RDG + 0.03°C |
Ni120 (= 672) | -40°C to +205°C | 0.01°C | 0.012% RDG + 0.03°C |
Ni1000 (= 618) | -60°C to +180°C | 0.01°C | 0.012% RDG + 0.03°C |
Cu10 (= 427) | -50°C to 150°C | 0.01°C | 0.012% RDG + 0.18°C |
Cu50 (= 428) | -50°C to +150°C | 0.01°C | 0.012% RDG + 0.06°C |
Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
Accuracies are given for 4-wire mounted probes
Take into account particular error of temperature sensor used and implementation conditions
Measuring current: 0.65 mA
Temperature coefficient: < 10% of accuracy /°C
Further features
Calibrated sensors | A database can be created to design curves for sensor s after calibration according to the corrections mentioned on a calibration report. |
Data recording | Data are recorded either manually on event or automatically with programmable frequency. Data is stored with date and time and can be displayed as list or curve. |
Statistical functions | Continuous display of average, minimum and maximum value of the signal under monitoring, as well as number of measurements. |
General specifications
Size | 157 x 85 x 45 mm |
Weight | 306 g |
Display | 160 x 160 pixel liquid crystal graphical display with backlite Display of result as table of values or trend curve |
Power supply | 4 AA batteries 1.5 V or rechargeable Ni-Mh batteries with internal charger in option |
Communication ports | USB |
Storage capacity | 10,000 data with date and time into one or several acquisition bursts |
Environmental specifications
Reference range | 23°C ± 5°C (RH: 45 to 75 % w/o condensing) |
Operating reference range | -10 to 50°C (RH: 20 to 80 % w/o condensing) |
Limit operating range | -15°C to +55°C (RH: 10 to 80 % w/o condensing) (70% at 55°C) |
Storage temperature limits | -30°C to +60°C |
Maximum height | 0 to 2,200 m |
IP protection | IP54 according to EN60529 |
Safety specifications
Protections |
|
Class | In accordance with EN 61010-1 Category II, pollution 2 |
Rated voltage | 60 V |
Chocks and vibrations | EN 61010-1 |
EMC conformity | Immunity:
EN 61000-4-4
|