제품소개 Products
PRIMARY METAL FREEZE POINT CELLS
PRIMARY METAL FREEZE POINT CELLS
PRIMARY METAL FREEZE POINT CELLS
PRIMARY METAL FREEZE POINT CELLS 모델명: K23

► Unique graphite "springs" allow the crucible to expand

    and the thermowell to float without risk of cell breakage

    inherent in other cell designs. When heated from room

    temperature to operating temperature, the graphite crucible

    and thermowell can expand as much as 1/8".

► The springs safely absorb this expansion while maintaining

    proper positioning of the crucible and graphite thermowell

제품 카달로그
견적문의
제품특징
상세사양

Pond Engineering's Primary Fixed Point Cells (Indium, Tin, Zinc, Aluminum and Silver on the ITS-90) combine outstanding precision, repeatability and accuracy with superb durability and ease of use to produce a uniquely convenient, reliable cell. Following the cell geometry described in NIST Technical Note 1265, these cells contain approximately 1.2kg of high purity metal (0.4kg for Aluminum and 1.6kg for Silver) in a machined graphite crucible with a graphite thermowell liner sealed within a quartz glass envelope. 

Unique to Pond Engineering's cells is a set of graphite "springs" directly above the crucible. These springs allow the thermowell to float and the crucible to expand more than 1/8" without stressing the envelope.  Two glass spacers above the crucible and thermowell liner maintain proper relative positioning of the thermowell liner.   Without this precaution, mechanical forces or hydrostatic pressures developed during heating can easily rupture the crucible, thermowell liner, or glass envelope; destroying both cell and thermometer.  

The outer envelope extends approximately four inches above the sealed portion of the cell and is fitted with several layers of ceramic fiber insulation to minimize axial heat loss and keep the cell top cool.  A machined aluminum cell top bonded to the envelope enhances durability, suspends the cell in the furnace and accommodates a removable handle.  These features ensure safe, easy cell insertion, removal and inspection - particularly helpful when the cell must be removed from the furnace to initiate a freeze (as with the Tin cell).

Since the defined Freeze Point conditions can only exist at the liquid-to-gas interfaces under one standard atmosphere pressure, our cells are sealed with high purity Argon at one standard atmosphere with the cell at operating temperature. Hydrostatic head corrections applied to the theoretical freezing point temperature compensate for the pressure of the liquid metal at the effective sensing point of the thermometer used in the cell. A user manual included with each cell includes tables and charts characterizing a typical freeze plateau profile and the cell's hydrostatic head correction.  The bound report also includes a certificate of metal sample purity. 

Metal Sample Purity > 0.999 999
Typical Melt Duration > 10 hours in
Pond Engineering
maintenance system
Cell Body Physical Dimensions ≈1.97" (50 mm) dia.
≈18.0" (46 cm) long
Thermowell Dimensions Quartz Glass
≈0.315" (8.0 mm) dia.
≈16" (406 mm) deep
Immersion Depth ≈7.5" (190 mm)
from liquid metal surface to well bottom
≈8.4" (214 mm)
from crucible top to well bottom
목록으로