제품소개 Products




독성 및 산소 결핍 가스 검지기 모델명: Sensepoint XCD
► Flammable, toxic and oxygen deficiency gas detector using catalytic,
electrochemical cell and infrared sensors.
► Used to monitor areas where flammable, oxygen deficiency and
toxic gases may pose a hazard to working environments.
► XCD offers high performance and reliability synonymous.
제품 카달로그
견적문의
Common transmitter platform
Same transmitter structure applicable to a variety of new and existing gas sensors
Local/remote sensor capability
Uses ECC/Catalytic/IR sensing technology
Smart intelligence recognises sensor type & range
Clear and informative display
Tri-colour backlight on LCD display
Instantly and clearly visible from a distance
Helps to clarify operation status in adverse situations
Multiple signal outputs
4 to 20mA standard
Switching relays
Modbus RTU
SurecellTM and ReflexTM
Superior performance sensing technology
Increased accuracy and reliability
Determines the health of O2 and toxic cells
Detects “dry-out” of the sensor cell
Ensures highest level of “up-time”
Infrared sensing technology
Low cost IR solution
Longer service life
No poisons
Doesn’t need oxygen
Non-intrusive operation
Easy routine maintenance
No need to open the transmitter
Faster planned maintenance
Greater operator confidence
Housing material options
LM25 Aluminium Alloy
Stainless Steel 316
Easy to install
Simple plug-in sensors
Integral plate for wall mount
Separate kit for vertical/horizontal pipe mount
On board relays allow for local alarms
3 wire operation, source or sink
Option remote sensor mounting
Option Modbus communications for remote diagnostics/Configuration
Worldwide certifications
IEC, ATEX
UL, CSA
KTL, GB
Sensepoint XCD Transmitter | |
Use |
3-wire, 4-20mA, gas detector transmitter for use with directly installed flammable and toxic gas sensors. For the protection of personnel and plant from flammable and toxic gas hazards. |
Electrical | Input Voltage Range : STD&RFD: 16 to 32Vdc (24Vdc nominal), RTD: 18 to 32Vdc (24Vdc nominal) |
Power Consumption : ECC = 3.7W, IR = 3.7W, Cat = 4.9W | |
Current output : 4-20mA (Source or Sink) | |
≥0.0<1.0 mA : Fault | |
4.0 mA to 20.0 mA : Normal gas measurement | |
2.0 mA or 4.0 mA (17.4 mA) : Inhibit (during configuration/user settings) | |
22.0 mA : Maximum over range | |
Terminals : 15 (STD), 18(RTD&RFD) x screw terminals suitable for wire diameter 0.5mm2 (20AWG) to 2.5mm2 (13AWG) | |
Relays : 3 x 5A@250VAC. Selectable normally open or normally closed (Terminal) and energized/de-energized (programmable) | |
Communication : RS485 Modbus RTU | |
Construction | |
Transmitter Material | Epoxy painted LM25 Aluminium. Stainless Steel 316 |
Sensor Material | Stainless Steel 316 |
Weight | LM25 Aluminium Alloy : 2.4kg. Stainless Steel 316: 4.5 kg |
Mounting | Pole or wall mounting |
Entries | 2 x M20 (or 2 x ¾”NPT) field cable entries, 1 x M25 sensor entry |
Performance | |
Repeatability | +/- 1% of FSD |
Linearity | +/- 1% of FSD |
Accuracy | < +/- 5% of FSD (O2, CO, H2S, FL-CAT) |
Response Time | T50 = <15s, T90= <30s (FL CAT and IR sensors) |
Detectable Gases |